Commercially Available Pacing Algorithms for Single-Lead Left Ventricular Multipoint Pacing and Possible Pacing Configurations
Feature Name | Boston Scientific MultiSite Pacing | St. Jude Multipoint Pacing | Medtronic Multiple Point Pacing | Biotronik MultiPole Pacing |
---|---|---|---|---|
CE Mark Status | Established 1Q 2017 | Approved | Approved | Approved |
Pacing Vectors Available | 17 | 10 | 5 | 12 |
LVa - LVb Timing Offset | Independent cathodes 0–100 ms | Independent cathodes 5–80 ms offsets | Tied cathodes, no offsets | Independent cathodes 0–50 ms |
Automatic Programming Recommendation | Yes SmartVector algorithm automatically recommends settings | Yes Options for choosing based on RV-LV or widest spacing | None for MPP | Unknown |
Pacing Configurations | Bi-V LV only | Bi-V | Bi-V LV only | Bi-V |
CE = European Commission; Bi-V = biventricular; LV = left ventricular; MPP = MultiPoint Pacing; RV = right ventricular.